Posted Wed, 23 Nov 2022 15:57:24 GMT by Grudovik, Ilya
The digitize 100V range behaves very strangely.
For example, I see a reaction from a change in input resistance.

Short input noise:
At 10 MΩ - 750 μV
At Auto - 65 μV

Input band:
At 10 MΩ - 17.5 kHz
At Auto - 7.5 kHz

Is there any difference between 10 MΩ and Auto?
If you select a range of 1000V there is no such effect.
Posted Wed, 07 Dec 2022 20:09:59 GMT by McKinney, Ty
Hello,
This would be a question and case best handeled by our technical support center. Would you create a case at the link below and reference this forum post? We should reach out to you once we have your information!
Posted Wed, 07 Dec 2022 20:20:20 GMT by McKinney, Ty
Support | Tektronix
Posted Thu, 08 Dec 2022 09:16:43 GMT by Grudovik, Ilya
I created.

You must be signed in to post in this forum.