Posted Tue, 13 Dec 2022 22:31:59 GMT by Kay, Jon
I have a TDS 2024B that I am trying to run through the adjustment procedure.  I'm using a Fluke 9500B O'scope Calibrator.  It runs through the DC voltages on channels 1 to 4 but fails when I get to the -2.5V step on EXT TRIG.  This is step 33 of the procedure.  Anybody have a possible solution or cause for this failure?  I would appreciate any ideas.

Thank you very much.
Posted Wed, 14 Dec 2022 00:18:38 GMT by Teles, Afonso
Hi,

The TDS2000B service manual lists step 33 (page 5-7) as 1.0 V, not -2.5 V.
Posted Wed, 14 Dec 2022 15:39:44 GMT by Kay, Jon
Thank you very much.... I had the manuel for the TPS model not the TDS.....very embarassing.

Thanks again.