Posted Wed, 15 Mar 2023 20:13:43 GMT by Parari, Vijaya Kumar
When I run Self Calibration  on TDS1102B, it fails at Step 1. If the selfcalibration fails at step 1 could any one help what it indicates (Bad reference voltage  ?). I am interested in knowing the possible rootcauses for this failure. Whether self calibration error could be corrected just by doing the factory calibration ? OR does the oscilloscope needs any hardware repair ?
Posted Fri, 17 Mar 2023 22:04:00 GMT by Teles, Afonso
Hi Vijay,<br> <br> I believe you had your question answered in a case but I will also answer it here so it may be useful to others:<br> <br> The self-tests test the front-end of the oscilloscopes, any failure here will require a hardware repair. Please make sure any signals and cables have been disconnected from the instrument before running the self-tests.<br> To set up a repair go to: <a href="https://www.tek.com/en/service-quote">https://www.tek.com/en/service-quote</a>